Description
SAVe II+ Ventilator Kit – Tactical Combat Casualty Care Configuration
Compact, economical, and easy-to-use, the SAVe II+ is built for dependable use during a crisis. Medical personnel have relied on AutoMedx products devices for over a decade. The new SAVe II+ delivers up to 50% more flow than the SAVe II and can generate up to 20 cmH2O of PEEP. This configuration includes a bladder face mask, head strap and syringe. The SAVe II+ replaces the SAVe II.
NSN: 6515-01-699-7669
This kit includes the following:
- 1x SAVe II+ Ventilator
- 1x Hard Case (water and dust proof), PN F20065
- 1x Power Charger, PN M42090
- 1x Patient Circuit, PN F20066 (3 year shelf life)
- 1x Extendable Oxygen Reservor Tubing, PN F20072
- 1x Mask (inflatible), PN E11000
- 1x Syringe to inflate mask
- 1x Head Strap, PN E11001
- 1x Noise Attenuator, PN M41112
General
Dimensions Ventilator 6.5″ x 6.25″ x 2.0″ (81 cubic inches)
Weight Ventilator 2.8 lbs. (1.3 kg)
Kit 9.9 lbs. (4.5 kg)
Operating ModeA/C-CMV
ModeTime cycled, pressure limited, volume targeted
PRESET Heights 4′ 3″ – 6′ 3″ in 3 inch increments
PRESET Target6 ml/kg of IBW or 30 cmH2O of Working Pressure
Flow Rate 0 – 40 LPM
I:E Ratio 1:2 Fixed
Respiratory Rate8 – 30 BPM
Tidal Volume 200 – 800 ml
Peak Inspiratory Pressure 10 – 60 cmH2O
PEEP 0 – 20 cmH2O
Inadvertent PEEP <2.0 cmH2O
FiO2 Range:21 – 100%
Oxygen Input:0 – 12 LPM from Flow Regulated Source
Inspiratory Trigger2.0 cmH2O
Inspiratory Resistance<6 cmH2O @ 30 LPM
Expiratory Resistance <6 cmH2O @ 30 LPM
Dead Space<100 ml
Battery Duration8.25-9.25 Hours (see below note)
Power Input100 – 240 VAC / 50-60 Hz
Audible Alarms Meets60601-1-8
Shock & Vibration IEC 60068-2-6 and 60068-2-36
Ingress Protection VentIP24
Protective CaseIP67
SETTINGS
Breath Rate (BPM)8 – 30
Tidal Volume (ml) 200 – 800
Pressure Limit (cmH2O)10 – 60
PEEP (cmH2O) 0 – 20
MEASUREMENTS
PIP (cmH2O) 0 – 60
PEEP (cmH2O) 0 – 25
FIXED PARAMETERS
I:E Ratio1:2
Inspiratory Time (Ti)0.67 – 2.50 seconds
Expiratory Time (Te)1.3 – 5.0 seconds
Flow Rate (Avg LPM)4 – 40
Product Life
Control Unit8 years
Breathing Circuit3 year shelf life (single use)
Notes
1. All specifications include a tolerance of ±10% of nominal value unless stated otherwise in the Operator’s
Manual. Test conditions available upon request.
2. Delivered Tidal Volume may be materially affected by very low lung compliance (<0.01 L/cmH2O)
3. With a fully charged new battery the 5’9” preset will run approximately 8.25 hrs with (2600 mAh battery) or 9.25 hrs with 2800 mAh battery)